Electronics Stocks
Description
?
?
Description:
?
Analytical binary high pressure gradient HPLC System
This binary system consists of two high pressure solvent delivery pumps, an UV/VIS detector,
a manual injection valve, and a solvent tray, a column and a Chromatography Workstation.
It is very simple for the users to install and operate this system. This system is very popular in normal lab now.
?
Ordering Details:
- - - -
Item | PN | Product description | QTY | Unit |
LC 3050 Analytical Binary high pressure gradient HPLC SYSTEM | 1 | set |
1 | L.F.P3005 | P3000 HPLC pump,10ml pump head ss,analytical | 2 | pcs |
2 | L.F.UV3023 | UV 3000?? UV Detector(Semi-Prep),with 50ml flow cell | 1 | pcs |
3 | I.IDEX.001 | 7725I Rheodyne Manual Injection Valve (20ul sample loop) | 1 | pcs |
4 | I.K.P.A0285 | Dynamic mixing Chamber ,Analytical and Semi-Prep | 1 | pcs |
5 | L.F.S.002B | Start-up Kit for Binary Analytical System | 1 | pcs |
6 | A.O.007 | Test column C18,4.6*250,5um,analytical | 1 | pcs |
7 | L.F.R3000 | Solvent tray,analytical | 1 | pcs |
8 | A.G.103 | CXTH 3000 Chromatography Workstation | 1 | pcs |
- - - -
?
Note:
* When the customers use the HPLC system for the first time, we suggest that solvent filter,
vacuum pump, ultrasonic cleaner, AC regulated power supply should be equipped.
?
?
Applications:
?
Pharmaceutical analysis: effective constituent analysis, related substance analysis
Food nutrients analysis: amino acid, sugar, organic acid etc.
Food safety (food contaminants and illegal additives etc.): melamine, Sudan, pesticide residue, plasticizer and aflatoxins etc.
Environmental pollutants analysis: PAHs, phenols and amine etc.
Chemical products and intermediates purity analysis
?
?
Technical parameter:
?
# P3000 High-Pressure Infusion Pump
Technical Parameters
Flow Rate Range: ( 0 - 9.999 ) mL/min
Max Working Pressure: 42 MPa
Flow Accuracy: ± 0.5 % ( 1 mL/min)
Flow Repeatability: RSD ≤ 0.1 % ( 1 mL/min )
PSU: 220 V ± 22 V, 50 Hz, 120 W
Dimensions and Weights: 450 × 270 × 180 mm( L × W × H ) 12.8 kg
?
# UV3000 UV-VIS Detector
Wavelength Range: ( 190 - 700 ) nm
Light Source: D2 lamp (standard configuration), Tungsten lamp (optional available)
Band Width: 8 nm, Wavelength accuracy: ± 2nm, Wavelength repeatability: 0.4nm
Baseline Noise: 1.5 × 10-5 AU/h
Baseline Drift: 4×10-4 AU/h
Dynamic: Velocity of flow 1mL/min; Methanol, wavelength 254 nm
Detection Cell Volume: 10 uL
PSU: 220 V ± 22 V, 50 Hz, 150 W
Dimensions and Weights: 450 × 270 × 180 mm ( L × W × H ) 10.4 kg
?
?
Competitive Advantage:
?
# Easy to adjust the gradient
# Easy to compose the modular units
# Positive and negative phase switch freely
# Provide system compatibility analysis of biological
Read More
Beijing chuangxintongheng Science &Technology Co., Ltd.
Preparative HPLC System,Industrial HPLC,Analytical HPLC System
Address: Rm401,Huizhong Building,Shangdi 7 Street,Haidian Dsitrict,Beijing,China,
Beijing, Beijing
China, 100000
Tel: 86-010-82890257
Fax: 86-010-82899729