We have two type, with filter and with grating. The one with grating is widely accepted in our market, below is its specifications:
Features
Open system design, free choice for reagent
Grating as dispersive element
Main specifications
Light source: Halogen-tungsten lamp (12 V/30W)
Wavelength range(nm): 330-800nm (continuous adjustment)
Bandwidth: ≤6nm
Absorbance range: -0.3Abs-3.0Abs
Peltier device: 25, 30, 37
Sample volume: 100ul--2000ul
Testing method:
A)Linear end point
B)Non-linear end point (standard curve)
C)Two-point (fixed time)
D)Kinetic
E)Single and double wavelength
F)Absorbance
G)Sample blank
Fixedly store 67 items, up to 200 user-defined items
Display: 240 X 128 LCD
Printer: Built-in thermal printer
Flowcell: 32ul; 10mm
Standard: RS-232 interface
Power supply: 220V 50Hz