Instruments & Meters
Description
The LED Optical Maintance and Life Test System (LEDLM-80PL) is according to IES LM-79, IES LM-80, IES LM-82, TM-12, GB2312-80, CIE and IEC Standards.
Refer to LM-80, the LED typically exhibit very long operational life characteristics and, depending on drive current and use conditions, can be in use for 50,000 hours or longer. Like all light sources, the light output from LED slowly decreases in output over time. Unlike traditional light sources, LED does not tend to fail catastrophically. Therefore, over time, lumen maintenance can result in lower light output than intended in the specification or required by codes, standard practices or regulations.
System Principle
Rated Lumen Maintenance Life, the elapsed operating time over which the LED light source will maintain the percentage, p, of its initial light output e.g:
?L70 (hours): Time to 70% lumen maintenance
?L50 (hours): Time to 50% lumen maintenance
The LED life is predicted according to the Arrhenius model, the formula as following:
1. P=P0 exp(-βt)
2. β=β0 IFexp(-Ea/KTj)
Note: P0 is initial lumen; P is lumen under heat and electricity; β is recession coefficient in a certain temperature; t is the time to electricity in a certain temperature; β0 is constant value; Ea is activation energy; K is Boltzmann& acute;s constant; IF is working current; Tj is junction temperature.
Firstly, we predict the LED life per 10000hours.
For example, we suppose the optical decay rate is n% in 1000hours, then:
Base on formula 1, L50 is: t=1000*ln0.5/ln(1-n%)
Base on formula 1, L30 is: t=1000*ln0.7/ln(1-n%)
Read More
Lisun Electronics Inc.
goniophotometer
Address: Room 1210,NO.9 Building,Jiangqiao Wanda Plaza,Hexuan Road,Jiading District,Shanghai,201803,China,
Jiading District, Shanghai
China, 201803
Tel: 0086-21-51083341
Fax: 0086-21-51083342