Others
Description
Product Details
Basic Info.
Model NO.:
UN700A
Measure Precision:
Medical Class
Normal Measure Time:
30s
Weak Signal:
10%
Operating Life:
100,000 Measurements
Export Markets:
Global
Additional Info.
Trademark:
UN
Packing:
Carton
Origin:
China
HS Code:
8534009000
Production Capacity:
20 000PCS/Year
Product Description
NIBP UN700A Specification:
1. Three Patient Mode, include Adult/PED
2. Test speed is 30s
3. Anti-motion
4. Continuous measurement mode
5. Calibration mode
6. Leakage check
7. Measurement accuracy reaches medical grade
8. Support bluetooth
9. Blood pressure measuring range 40 ~ 255mmHg
10. Weak signal test (CUFFLINK simulator test) is 10%
Appendix A Performance specification
1.Performance specification
1.1 Measuring range
Adult:
SYS: 40 ~ 255mmHg
MEAN: 20 ~ 215mmHg
DIA: 10 ~ 195mmHg
PULSE RATE: 40 ~ 240bpm
PED:
SYS: 40 ~ 200mmHg
MEAN: 20 ~ 165mmHg
DIA: 10 ~ 150mmHg
PULSE RATE: 40 ~ 240bpm
1.2 Measurement precision
blood pressure precision :
Average deviation: <=±5mmHg, standard deviation: <=8mmHg
Pulse rate precision
Max. deviation: <=2% (>=100 BPM) or 2 beats/min (<100 BPM)
1.3 software protect pressure
Adult :300mmHg
Ped:240mmHg
1.4 Pressure precision and resolution
Pressure precision
+/-3mmHg
resolution
1mmHg
1.5 Operating environment
Operating temperature 0~45oC
Operating humidity 15%~90%, non-condensing
Storage Temperature -20~60oC
Storage humidity 0 ~ 95 %, non-condensing
Operating altitude -500 meters ~ 5000 meters
Operating life 100,000 measurements